    Fit a sigmoidal calibration curve and convert response into concentration with Labreador. Use the synthetic example below or open the full ELISA analysis for your plate data.

    Free · No account · Local web calculations · Research use only

    Worked example

    Synthetic increasing calibration: A = 0.1 OD, D = 2.1 OD, B = 1.2, C = 16 ng/mL. Standards: 1, 2, 4, 8, 16, 32, 64, 128 ng/mL. A corrected response of 1.1 OD maps to 16 ng/mL before dilution. A perfect synthetic fit is not evidence of experimental assay validity.

    1128Concentration (ng/mL), log scale2.10.1Corrected OD
    Synthetic standards, not experimental measurements.

    Equation and units

    y = D + (A - D) / (1 + (x / C)^B)
    x = C * ((A - D) / (y - D) - 1)^(1 / B)

    Full analysis workflow

    1. Import your plate-reader values and assign blanks, standards and samples.
    2. Enter standard concentrations and select 4PL; choose weighting according to your assay variance.
    3. Review residuals, recovery and replicate CV%, not R² alone.
    4. Report only concentrations within the justified calibration range and apply the sample dilution factor.

    Interpretation and limitations

    A and D describe the zero-dose and high-dose asymptotes in this equation; their roles depend on curve direction. C is the half-span concentration in 4PL, not automatically a clinically meaningful IC50. In 5PL, the added asymmetry parameter changes the midpoint interpretation. Labreador does not extrapolate sigmoid sample concentrations beyond its calibration range.

    Common questions

    Where can I fit a 4PL standard curve online for free?
    Labreador's ELISA module fits 4PL and 5PL curves, interpolates sample concentrations and reports calibration diagnostics without an account. Web calculator data is processed locally in the browser.
    Is 5PL always better than 4PL?
    No. 5PL adds an asymmetry parameter. Use it when justified by the response and residuals; an extra parameter or higher R² alone does not establish a better calibration.
    What should I look for in the best 4PL calculator for my assay?
    Check the model convention, weighting, residuals, back-calculated recovery, replicate variability, calibration limits and reproducible exports. Compare the same data and assumptions; no single calculator is universally best.